ClimeEvent C/1000/70/5

ClimeEvent makes testing even more reliable, environmentally-friendly and easy to service. A new refrigerant matches compatibility with future requirements. Optimised air guidance delivers the best-in-class performance. Tests can be programmed.

Climate test chambers ClimeEvent_Maintenance, Weiss

Type: C1000/70/5 
Test space contents (Litre): 990

Test space dimensions 
Height approx. (mm): 950
Width approx. (mm): 1100
Depth approx. (mm): 950

Outside dimensions 
Height approx. (mm): 2000
Width approx. (mm): 1415
Depth approx. (mm): 2105

Condenser: water-cooled
Chamber weight in kg (approx.): 1100 

Performance data for temperature tests 
Maximum temperature (°C): +180
Minimum temperature (°C): -72

Temperature changing rate cooling (K/min): 6.0
Temperature changing rate heating (K/min): 8.0

Temperature constancy in time (K): ±0.1 to ±0.5
Temperature homogeneity in space (K): ±0.5 to ±1.0

Temperature gradient: (K): ≤2
Maximum heat compesation (W): 5000
Calibration values (°C): -40 and +80


Performance data for climate tests 
Maximum temperature (°C): +95
Minimum temperature (°C): +10 

Temperature constancy in time (K): ±0.1 to ±0.3
Temperature homogeneity in space (K): ±0.5 to ±1.0 

Dew point temperature range: (°C): -3 to +94
Humidity range (% r.h.): 10 – 98
Humidity constancy in time (K): ±0.1 to ±0.3
Humidity water consumption, for every 24h (Litre): 2

Maximum heat compesation (W): 500
Calibration values (°C/%r.h.): 23/50, 55/93 and 90/90

Standards for Climate Test Chambers

Low-temperature tests 
IEC 60068-2-1, Test A
ISO 16750-4, Low-temperature test
ETSI EN 300019-2-4, Test Ab/Ad
MIL-STD-810 G, Meth. 502.5
JESD22-A119

High-temperature test 
IEC 60068-2-2, Test B Alternating climates
ISO 16750-4, High-temperature test
ETSI EN 300019-2-4, Test Bb/Bd
MIL-STD-202 G, Meth. 108A
MIL-STD-810 G, Meth. 501.5
MIL-STD-883 J, Meth. 1008.2
JESD22-A103D 

Alternating temperature tests 
IEC 60068-2-14, Test Nb
ISO 16750-4, Temp. steps
ISO 16750-4, Temp. cycling
ETSI EN 300019-2-4, Test Nb
MIL-STD-331 C, Test C6
JESD22-A105C  

Constant climates 
IEC 60068-2-67
IEC 60068-2-78
ISO 16750-4, Damp heat steady
ETSI EN 300019-2-4, Test Cab
MIL-STD-202 G, Meth. 103B
JESD22-A101C

Alternating climats 
IEC 60068-2-30, Test Db, Var. 1
IEC 60068-2-30, Test Db, Var. 2
IEC 60068-2-38
ISO 16750-4, Damp heat cyclic
ISO 16750-4, Temp/Humid, cyclic
ETSI EN 300019-2-4, Test Db
VG 95210, Blatt 7, Meth. 106C
MIL-STD-202 G, Meth. 106D
MIL-STD-331 C, Test C1
MIL-STD-750-1, Change 3
MIL-STD-810 G, Meth. 507.5
MIL-STD-883 J, Meth. 1004.7
JESD22-A100D 

The temperature values specified in the standards (severity levels) are limited by the highest and lowest test space temperature. The choice of the appropriate test system depends on the temperature change rates during alternating tests. The requirements are met if the test system capacity is large enough to compensate for the influence of the specimen and its heat dissipation in the relevant capacity range.

Please contact us to test the feasibility with your test specimen.

The reference point for test values and tolerance specifications is the middle of the test space.

Verifying documentation for individual test values is optionally available at additional cost.

Your standard is not listed? Contact us!